A Toe Brachial Index or TBI is performed when the ABI or Ankle Brachial Index is abnormally high due to plaque and calcification of the arteries in the leg; this is caused by atherosclerosis and is most often found in diabetic patients. The abnormally high ABI is >1.3.
Equipment Needed:
The TBI Package which includes the MD6RP Photo plethysmograph and photo transducer and digit cuffs (two UPC 2.5) as well as the Portable ABI Kit which provides the MD6VR Chart Recorder, S400 Aneroid Sphygmomanometer, and two SC12 and 2 SC10 Straight Segmental Cuffs.
Preparation of the Patient:
The patient should be in a warm comfortable room, supine with arms and legs at heart level. Keep feet warm with a blanket or towel if needed.
The Procedure:
Select the appropriate sized cuff for each limb. You will need one cuff for each upper arm and for each large toe. Measure the cuff width to the diameter of the limb; the cuff width should be 20% larger than the limb diameter to compress all of the soft tissue evenly. The cuff should be put on straight and fit snuggly but not tight. You must use vascular cuffs which have long bladders to completely encircle the limb and compress all of the soft tissue. The toe cuff should be wide enough to apply pressure over a large enough area so as to not be a tourniquet and long enough to over lap the bladder.
Put the cuffs on the arms and toes once the patient is supine. Have the sphygmomanometer, patient chart, etc. in the room ready to use. A patient can relax best if undisturbed.
The rest period should be at least 10 minutes with 15-20 ideal. The goal is for the patient to relax. This time can be used for interviewing the patient, listening to the heart, palpating the abdomen, checking the neurological reflexes and sensation in the feet.
After the rest period, take the first brachial pressure. Find the brachial pulse with your fingers, and then put some ultra phonic gel on that place. Obtain a good Doppler signal (sound) and waveform, print out to establish a base line. Inflate the cuff until the sound and waveform disappear and then inflate from 20-30 mmHg above that number (super systolic). Slowly deflate the cuff (around 2-3 mmHg at a time) until the sound reappears (Korotkoff sound) the waveform will follow immediately after the sound. The pressure reading when the first sound appears is the correct systolic pressure. Deflate the cuff completely and record the systolic pressure.
Toe Pressure Demonstration