Assay: Accuracy and Precision with Serial Dilution (2024)

March 15, 2007 (Vol. 27, No. 6)

Many laboratory protocols require the serial dilution of reagents or compounds. IC50 assays, commonly used to evaluate drug efficacy, and assay development procedures, as well as standard-curve generation, involve the serial dilution of compounds, proteins, or detection agents. These processes can be streamlined by utilizing automated liquid-handling equipment with serial dilution capabilities.

Serial dilution processes face two major challenges. The first is error propagation across columns or rows. With each sequential serial dilution step, transfer inaccuracies lead to less accurate and less precise dispensing. The result is that the highest dilutions will have the most inaccurate results. To compensate for this error possibility, longer mixing times are required, which then increases the time required to perform the serial dilution. These challenges greatly limit the throughput capacity of an automated serial dilution system.

To overcome these challenges, the effects of various mixing parameters of a serial dilution protocol were explored. Velocity11’s (www.velocity11.com) Bravo™ Liquid Handling Platform performed serial dilution with the same pipette head as a full plate dispenser (Figure 1). With the platform’s VWorks™ software, the application allowed the total control of liquid transfer and mixing heights and speeds, which allowed efficient exploration of mixing parameters. The goals were to determine which parameters had the greatest effect on mixing and to reduce the time required to perform a serial dilution.

Assay: Accuracy and Precision with Serial Dilution (1)Serial Dilution Mix Cycles

The basic experiment diluted fluorescein across the columns of a 96-well plate, from A1 to A10 (A11 and A12 were blank wells). The starting volume was 300 µL, and 200 µL tips were utilized for the transfer (150 µL, a 1:2 dilution) and mixing steps (190 µL). There are two main components of an accurate and precise serial dilution: the accuracy and precision of the transfer and the efficiency of mixing. Transfers were previously determined to have a precision and accuracy of >99% at this volume; any observed deviations in precision and accuracy were due to error propagation from ineffective mixing.

Two measures were used to evaluate mixing efficiency. The Coefficient of Variance (CV) of each column indicated the precision of the mixing step. The CV also provided information on the propagation of error across a plate—the CV increased sharply across the plate if mixing was incomplete.

The second indication was the accuracy of the transfer. A calibration curve was prepared, and each experimental dilution concentration was plotted against the standard curve to determine the real concentrations in each column. The first experiment varied the number of mixing cycles between 3 and 20. The average precision (averaging CVs for columns 1–10) improved asymptotically as the number of mix cycles increased. Three mixes before each transfer yielded an average CV of 11.8%, while 20 mixes gave a considerably better CV of 1.7%.

The precision in all cases generally worsened as the serial dilution proceeded across the plate; this was expected as the error in the earlier columns propagated with each transfer.

The accuracy ratio improved as the number of mix cycles increased. The accuracy ratio is an average of the concentration of the diluted column compared to the previous column—a perfect serial dilution has an accuracy ratio of 1:2.00 across the entire plate. The accuracy ratio of the plate improved with more mix cycles, improving from 1:1.85 to 1:2.01.

While the precision and accuracy with 20 mix cycles is close to a perfect serial dilution, the length of time required might be considered impractical. The 20-mix cycle protocol required 20 minutes per plate, while a three-mix cycle protocol required less than six minutes. Efforts were then focused on the factors that could improve the three-mix cycle protocol to produce accuracy and precision results consistent with the 20-mix cycle protocol.

Mix Tip Height

The mix tip height was modified in order to determine the effect of distributing the liquid at different locations in the well. As the mix tip height was raised, the average precision improved. At a height of 3 mm from the bottom of the well, the average precision was 3.9%. The precision worsened as the tip distance from the bottom of the well decreased, reaching a CV of 15% at a height of 0.1 mm. Accuracy tracked with precision, and the higher mix height also improved the accuracy ratio to 1.95. This trend is possible because the higher dispense height ensures that more of the sample was circulated by the mix cycle.

In a mix roughly in the middle of the well volume, the dispensed liquid is forced toward the well bottom while dispensing, and aspirated liquid is pulled from the center of the well. If the mix occurs close to the bottom of the plate, the dispensed liquid is pulled back into the tip during the aspiration. Mixing in the center allows the dispensed liquid to be more evenly distributed in the sample, thus increasing the likelihood of efficient mixing.

Assay: Accuracy and Precision with Serial Dilution (2)

Mix Liquid Class Setting

The VWorks software controlling the Bravo platform allows the creation of liquid classes, which allows the operator to modify the velocity and acceleration for aspirating, dispensing, and mixing tasks. The original liquid class settings for the mix were 100 µL/s velocity and 500 µL/s2 acceleration. Precision and accuracy improved as the mix velocity increased. This effect plateaus; above 300 µL/s,

there is no appreciable improvement in increasing the speed. The cause of this is likely due to the creation of more turbulent mixing, which in turn distributed the fluorescein dye more quickly throughout the solution.

Dynamic Tip Retraction/Extension

Finally, the effect of dynamic tip retraction and extension was explored. This function moved the tips deeper intothe well during each aspirate step, and retracted them during each dispensing step. This allowed a larger volume of the well to be affected by the mix step by adding the movement of the tip into the mix task. There was a marginal improvement (less than 0.5% improvement in CV/accuracy) observed in using this technique.

Additionally, no effect was observed by utilizing another mix standard, which involved aspirating close to the bottom of the well and dispensing near the top of the solution. This mixing method caused no improvement once the other parameters described above had been optimized. These experiments mixed hom*ogenous solutions; there may be an improvement with this technique if the solutions are expected to have different viscosities.

Conclusion

Based on these experiments, the parameters that had the largest impact on efficient mixing were (in decreasing order):

  • Speed of the mixing step
  • Height of the tip during the mix
  • Tip-retraction capabilities

To verify this conclusion, the first experiment (varying the number of mix cycles) was repeated with the improved mix parameters. The new parameters provided increased precision and accuracy and improved the accuracy and precision of the 3-mix cycle operation to a level comparable with the 20-mix cycle operation (Figure 2). More importantly, the new parameters also decreased the time required to run an effective serial dilution protocol from 20 minutes to just under 5 minutes. This has tremendous potential in automating a serial dilution assay and ensuring accurate and precise results.

InsightsResourcesTutorialsAnimal modelsBacteriaCloningDisease modelsEscherichia coliFermentation byproductsFood productGenetic engineeringGenetic Engineering and BiotechnologyHealth care and economicsHealth care facility, manpower, and servicesHealth care organizationsIndustrial biotechnologyMicrobiologyMicroorganismsResearch and developmentUnited States Food and Drug Administration

Assay: Accuracy and Precision with Serial Dilution (2024)

FAQs

What is the accuracy of serial dilution? ›

The accuracy ratio is an average of the concentration of the diluted column compared to the previous column—a perfect serial dilution has an accuracy ratio of 1:2.00 across the entire plate. The accuracy ratio of the plate improved with more mix cycles, improving from 1:1.85 to 1:2.01.

Which gives more accurate results, serial dilutions or independent dilutions? ›

Dilution experiments are subject to both measurement and pipetting errors; the latter may cause systematic and accumulative errors in dilution. Gravimetric, independent dilutions are preferred to serial dilutions.

How do you make serial dilutions more accurate? ›

The accuracy and precision of serial dilutions also depend on a thorough mixing at each step: this is essential to achieve a hom*ogeneous mixture with a constant concentra- tion over the sample volume. Without a proper mixing, any ali- quot can have a higher or lower concentration than expected, randomly (Figure 1).

Is serial dilution more accurate than direct dilution? ›

So, preparing a concentrated stock solution that may require a serial dilution involving transfers of larger volumes will reduce the measurement uncertainty of the final solution concentration and, therefore yield a more accurate final concentration.

What can serial dilution tell you? ›

The purpose of a serial dilution is to estimate the concentration of a sample, or to obtain the desired concentration of a reagent, chemical or compound.

What can go wrong in serial dilution? ›

For example, incomplete mixing and human error, often compounded by problems such as a time constraint on the process, may contribute to the problem. It will be seen that errors of this magnitude can result in dramatic increases in the variation of the estimate of the initial concentration.

What is the main disadvantage of the serial dilution technique? ›

However, there are also disadvantages associated with serial dilution-agar plate procedure. The primary disadvantage is that the process is time-consuming and requires a high level of skill to perform accurately. Also, cultures grown on agar plates generally don't survive for long due to dehydration.

What is the main advantage of serial dilutions? ›

What is the purpose of serial dilution? Serial dilution is used to obtain the desired concentration by diluting the starting/highly concentrated solution with an appropriate liquid. It is often used to avoid transferring very small volumes to make a highly diluted solution.

Which dilution is most accurate? ›

When doing very high dilutions (like 1/10,000 or 1/1,000,000), it is more accurate to do the dilution in a series of smaller dilutions rather than in one giant dilution. This is called a dilution series or a serial dilution.

Why do we perform serial dilutions two times but get two different results? ›

Why do we perform serial dilutions two times, but get two different results? What makes the webbed plate different from the plate where we calculate the Titer PFU? We perform serial dilutions twice because we need a webbed plate to flood and a PFU Titer to perform calculations on.

Do you need to change tips for serial dilution? ›

There is no need to change GRIPTIPS during serial dilutions of aqueous solutions, as long as a blowout is performed after each mixing step to prevent liquid carryover.

How to improve serial dilution? ›

Mixing is essential for an accurate serial dilution. A mixing step using half the total volume of the intermediate step introduced after every transferring step ensures an accurate and consistent serial dilution.

What is the alternative to serial dilution? ›

Direct dilutions

The alternative approach to serial dilution is the direct dilution of micro-volumes of compound (ie on a volumetric basis).

Why a series of dilutions is a more accurate and or convenient method? ›

**Error Minimization**: When conducting multiple dilutions, any errors made in one dilution step can be corrected in subsequent steps. This helps to minimize the impact of errors and ensures a more accurate final concentration.

What is the advantage of using serial dilutions instead of simply weighing out a sample and using volumetric glassware to make it to the molarity you want? ›

No balance in existence can measure that tiny of an amount. But you can make a more concentrated solution and do several successive dilutions easily. It's the only way to make such a dilute solution accurately since you cannot do it by weighing such a tiny mass.

What is accurate dilution? ›

Accurate dilutions are one of the best used tools in microbiology for determining the number of microbes in something. No matter how many bacteria are in your initial sample, you can always dilute it down to a countable level.

What is the result of the serial dilution? ›

A serial dilution is the step-wise dilution of a substance in solution, either by using a constant dilution factor, or by using a variable factor between dilutions. If the dilution factor at each step is constant, this results in a geometric progression of the concentration in a logarithmic fashion.

What is the normal range for serial dilution? ›

Serial dilutions of samples are made using EMEM10 as the diluent in 8-well Lab-Tek slides. Typical dilutions are 1:2.5, 12.5, 62.5, and 312.5 in a final volume of 0.1 mL (a screening test can be performed by making two dilutions, e.g. 1:5 and 1:25).

Top Articles
What to Know About Split Tunneling - Intraprise Health
What is HPB? – Find out more about the Holiday Property Bond
Botw Royal Guard
³µ¿Â«»ÍÀÇ Ã¢½ÃÀÚ À̸¸±¸ ¸íÀÎ, ¹Ì±¹ Ķ¸®Æ÷´Ï¾Æ ÁøÃâ - ¿ù°£ÆÄ¿öÄÚ¸®¾Æ
Chelsea player who left on a free is now worth more than Palmer & Caicedo
Craigslist Free Stuff Appleton Wisconsin
Tv Schedule Today No Cable
The Haunted Drury Hotels of San Antonio’s Riverwalk
Tiraj Bòlèt Florida Soir
Uc Santa Cruz Events
Osrs Blessed Axe
Thayer Rasmussen Cause Of Death
Nonuclub
Miss America Voy Forum
charleston cars & trucks - by owner - craigslist
Jenn Pellegrino Photos
Equipamentos Hospitalares Diversos (Lote 98)
Abortion Bans Have Delayed Emergency Medical Care. In Georgia, Experts Say This Mother’s Death Was Preventable.
Ess.compass Associate Login
The Exorcist: Believer (2023) Showtimes
Shadbase Get Out Of Jail
Ihub Fnma Message Board
Ontdek Pearson support voor digitaal testen en scoren
§ 855 BGB - Besitzdiener - Gesetze
'Insidious: The Red Door': Release Date, Cast, Trailer, and What to Expect
Rgb Bird Flop
Ipcam Telegram Group
Desales Field Hockey Schedule
Trust/Family Bank Contingency Plan
How to Use Craigslist (with Pictures) - wikiHow
Pch Sunken Treasures
Quality Tire Denver City Texas
Powerball lottery winning numbers for Saturday, September 7. $112 million jackpot
A Man Called Otto Showtimes Near Amc Muncie 12
Tds Wifi Outage
How to Draw a Sailboat: 7 Steps (with Pictures) - wikiHow
Rochester Ny Missed Connections
Google Flights Orlando
Mid America Irish Dance Voy
Nba Props Covers
Isabella Duan Ahn Stanford
Trivago Anaheim California
LumiSpa iO Activating Cleanser kaufen | 19% Rabatt | NuSkin
John M. Oakey & Son Funeral Home And Crematory Obituaries
BCLJ July 19 2019 HTML Shawn Day Andrea Day Butler Pa Divorce
Tommy Bahama Restaurant Bar & Store The Woodlands Menu
Cara Corcione Obituary
Costner-Maloy Funeral Home Obituaries
antelope valley for sale "lancaster ca" - craigslist
Thrift Stores In Burlingame Ca
Basic requirements | UC Admissions
Guidance | GreenStar™ 3 2630 Display
Latest Posts
Article information

Author: Sen. Ignacio Ratke

Last Updated:

Views: 6231

Rating: 4.6 / 5 (56 voted)

Reviews: 87% of readers found this page helpful

Author information

Name: Sen. Ignacio Ratke

Birthday: 1999-05-27

Address: Apt. 171 8116 Bailey Via, Roberthaven, GA 58289

Phone: +2585395768220

Job: Lead Liaison

Hobby: Lockpicking, LARPing, Lego building, Lapidary, Macrame, Book restoration, Bodybuilding

Introduction: My name is Sen. Ignacio Ratke, I am a adventurous, zealous, outstanding, agreeable, precious, excited, gifted person who loves writing and wants to share my knowledge and understanding with you.